Section XV – Electromagnetic Conformity Declaration
The Kangaroo™ Joey enteral feed and flush pump with pole clamp has been built and tested according to UL 60601-1 (2nd
Edition), ES60601-1(3rd Edition), CAN/CSA C22.2 No. 60601-1-08, and EN60601-1-2 Standards.
The Kangaroo™ Joey enteral feed and flush pump with pole clamp is intended for use in the electromagnetic environment
specified below. The user of the pump should assure that it is used in such an environment.
The Kangaroo™ Joey pump is intended for use in the electromagnetic environment specified below. The user of the Kangaroo™ Joey pump
should assure that it is used in such an environment.
Emissions Test
RF emissions
(CISPR 11)
RF emissions
(CISPR 11)
Radiated Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-3: 2006 + A1: 2007 +
A2: 2010)
Conducted Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-6:2013)
Power Frequency Magnetic Field Immunity
(EN60601-1-2 / IEC 61000-4-8:2009)
Voltage dips and sags Immunity
(EN60601-1-2 / IEC 61000-4-11:2004)
Electrical Fast Transient / Bursts Immunity
(EN60601-1-2 / IEC 61000-4-4:2012)
Electrostatic Discharge Immunity
(EN60601-1-2 / IEC 61000-4-2:2008)
Surge Immunity
(EN60601-1-2 / IEC 61000-4-5:2005 +Cor 1: 2009)
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Compliance
Group 1
Class B
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Electromagnetic Environment - Guidance
The Kangaroo™ Joey enteral feed and flush pump with pole clamp
uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The Kangaroo™ Joey enteral feed and flush pump with pole
clamp is suitable for use in all establishments, including domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.