Operation; Machine Statuses; Load The Materials; Handpieces And Turbines - Mocom Millwashing Manuel D'utilisation

Masquer les pouces Voir aussi pour Millwashing:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

OPERATION

MACHINE
STATUSES
LOAD THE
MATERIALS
Handpieces and
turbines
The unit can take the following statuses:
STANDBY: The unit is ready to start a cycle.
Self diagnostics are active.
CYCLE:
The unit enters in this status with the Start command that is only accepted for
machine in standby and door closed.
The cycle starts and the various phases are successively executed. Self
diagnostics and controls are active. The display shows the current value of the
temperature.
BLOCK:
The internal diagnostics has detected an anomaly involving the block of the
operation; the cycle is interrupted and the door remains blocked. The anomaly
is signaled with the red Led (
door unblock procedure allowing the return in Standby state (see door
unblocking procedure).
Insert the instruments to be washed by carefully positioning them in the special holders
and tray.
-
The objects must not overlapped.
-
Position the objects so as to avoid deposit of water inside or on its external surface.
-
Arrange high and heavy objects in the centre of the basket to facilitate the washing.
ARRANGE THE OBJECTS SO THAT THEY DO NOT BLOCK THE
WASHING ARMS; CHECK THAT THEY CAN TURN FREELY.
T
HE MAXIMUM OVERALL LOAD ALLOWED FOR EACH CYCLE IS
The washing and thermo-
disinfection
of
hollow
instruments must take place
both outside and inside the
cavity. To this purpose, the
basket is equipped with
special
accessories
in
which the handpieces and
turbines can be plugged.
For the washing of pipes,
fittings of different diameter
are available.
W
ASH THESE FILTERS WEEKLY OR REPLACE THEM
The program for the correct washing and thermo-disinfection of straight and angled
handpieces and turbines includes:
washing at 65°C with neuter PH detergent and free of mineral substances.
-
-
thermo-disinfection at 93°C, temperature for 1 minute, with the addition of additive
to eliminate traces of water residue.
The suggested chemical products are:
-
MEDICLEAN, neuter PH detergent.
MEDIKLAR, additive for the elimination of water residue.
-
D
O NOT SUBMIT MICROENGINES TO THERMO
F
OR HANDPIECE AND TURBINE TREATMENT
N
. 1
PROGRAM
R
D
O NOT USE WASHING POWDERS
), and the machine is waiting for the
WARNING
NOTE
Hand-piece
Fitting
Pipe
to be whashed
NOTE
NOTE
-
DISINFECTION
,
IT IS RECOMMENDED TO USE ONLY THE
N
. 2
AND NOT THE PROGRAM
R
OR
.
8.0
.
KG
rubber adapter
ring nut
filter
The handpiece accessories
are made of a rubber
adapter (different diameters
available), ring nut and filter.
.
.
3.
11

Publicité

Table des Matières
loading

Table des Matières