V.1 – Connection to the motor
Check that the MD is completely dry before connecting it to the
motor.
Install the MD on the motor connection until ratcheting. For this,
keep the motor and the MD in the same axle. The motor must be
at a complete standstill.
Apply light traction to the MD to check that it is properly attached
to the motor connection.
Operational test by switching on the motor; start at low speed,
then gradually increase, outside of the patient's mouth.
If you notice overheating, irregularities, vibrations, abnormal
noises when using the MD, immediately contact your AS.
V.2 – Adjusting the declutching ring (Fig. 3)
The torque control system cannot in any way be a substitute for
the practitioner's assessment and experience.
Increasing the declutching threshold :
• Push ring G towards Head E on the MD, notch by notch, until the
right position is reached.
Decreasing the declutching threshold :
• Pull ring G towards the body H on the MD, notch by notch, until
the right position is reached.
To adjust the declutching threshold (position 1, 2, 3, 4), start by
setting it at its minimum torque (position 1).
If the MD keeps declutching, move up to a higher torque (Positions
2 to 4).
When the radicular instrument stops, a slight clicking sound may
be heard; in that case, lower the pressure on the MD.
In the event that the instrument becomes wedged in the root
canal, reverse direction of the motor rotation.
V.3 – Disconnection of motor
Remove the MD by keeping it in the motor axle.
D
In the event of prolonged non use of the device, do not
leave it connected to the motor. Risk of damaging the
motor.
V.4 – Connection/disconnection of rotary instrument
r
For all handling of rotary instruments, wear protective
gloves. Risk of cutting and contamination. The tool must
only be handled when the motor is at a standstill.
Only use rotary instruments in good condition. Risk of necrosis.
Insertion of rotary instrument
Contra angle with push button :
• Insert the rotary instrument into the head of the MD until it hits
the stop.
• Turn the rotary instument whilst maintaining a slight axial
pressure until it clicks (Fig. 4).
Contra angle with latch :
• Open the latch.
• Insert the rotary instrument into the head of the MD until it hits
the stop.
• Turn the rotary instument whilst maintaining a slight axial
pressure until it reaches the running slot on the drive plate.
• Lock the latch (Fig. 5).