Layout of the unit
1. copper rod
2. contact
3. distance piece
4. silicone seal
5. beaker
6. anode/temperature sensor
7. protective ring
8. front plate (with diagrams)
9. electroforming head mount
10. contacts for denture base and
residual gold deposition
11. electroforming head
12. crown coping
13. hotplate cover
Layout of the controls
14. power level button
15. LED display
16. holes for copper rods
17. status display
18. timer button
19. Start/Stop button
20. operation indicator (LED)
Buttons and displays
Power level button: for selecting the required power
level (can be adjusted individually
for each channel)
LED display:
indicates the power level set for
each channel
Timer button:
for selecting the programs:
1: electroforming thickness
approx. 0.2 mm
operating time 300 min
2: electroforming thickness
approx. 0.3 mm
operating time 415 min
3: residual gold deposition
operating time 120 min
4: lower denture base
operating time 390 minn
5: upper denture base
operating time 540 min
Status display:
Indicates the program set, the gold
electrolyte quantity, the remaining
electroforming time and errors
Start/Stop button: for starting and stopping the
electroforming process
!
Attention:
Do not start the unit without gold
electrolyte!
Operation indicator: Lights during the electroforming
process.
6