Varian Turbo-V 700 ICE Notice De Mode D'emploi page 88

Table des Matières

Publicité

Choose the desired selection by pressing the CYCLE
push-button.
F R O N T
/
R E M O T E
S E L E C T
I O N :
Press PUMP CURRENT push-button and the display
shows:
S E R I
A L
This message will only be displayed if the RS 485 module
is installed.
Where:
XX = 00 ÷ 31 is the unit address
/
S E R
F R O N T
A D D R E S S
X X
Press the PUMP CURRENT push-button to enter the
value and ,if the serial option is installed, the display
shows:
R S X X X
B A U D
R A T E
where: BAUD RATE XXXX = means 600, 1200, 2400,
4800, 9600 baud rate for the host computer or printer
communication.
The controller is factory-set for 9600 baud rate operation.
The value which corresponds to the board installed will be
assigned to RSXXX.
If the RS 232 is installed, enter the value by pressing
the PUMP CURRENT push-button and, if the serial
option is installed, the display shows:
R S X X X
S E L E C T
This message will only be displayed if the RS 232 interface
is present.
where:SELECTION: XXXXX = means HOST or PRINT.
The controller is factory-set to HOST
84
TECHNICAL INFORMATION
X X X X
H O S T
/
P R I
I O N :
X X X X X
87-900-904-01 (I)
N T