Agilent Technologies TwisTorr 84 FS PCB 24 Mode D'emploi page 182

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 33
17
Technical Information
Installation
The following figure shows a typical contact logic input connection
and the related simplified circuit of the controller.
Figure 10
182/210
Pin 11 Error reset command. Should an operation fault be
detected (Pin 3 of J4 is ON), the pump is stopped. To reset the
fault you must connect pin 11 to ground (pin 15 of J4) for at least
one sec.
Pin 12 ON/OFF pump. To stop the pump you must connect pin
12 to ground (pin 15 of J4); to start the pump you must remove
the contact between pin 12 and ground.
Pin 13 Low Speed mode activation. When the contact between
pin 13 and ground (pin 15 of J4) closes, the turbo pump runs at
low speed and when the contact opens, the turbo pump reverts to
high speed mode.
When the Low Speed mode is activated pin 5 of J4 is ON.
The Low Speed frequency is 1100 Hz.
Pin 15 Ground
Typical Logic Input
TwisTorr 84 FS PCB 24 Vdc Controller User Manual / 87-901-039-01 (B)
+5 V

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

X3510-64000

Table des Matières