PROBLEM SOLVING
IF THE EXTENSION SET TUBING BECOMES DISCONNECTED
FROM THE EXTENSION SET
1. Stop the pump.
2. Estimate the amount of formula lost.
3. Thoroughly wipe the tube connections with soap and water
or alcohol. They must be free from oil or formula build-up.
4. Clean inside the Feed Port with a cotton-tipped
applicator and alcohol.
5. Flush the extension set tubing with warm water.
6. Dry the connections and firmly reconnect
the extension set (See
Right Angle connector is fully seated. (See
7. Resume the feeding, replacing the estimated volume
lost during the disconnection.

BALLOON LEAKS OR RUPTURES

Always keep a replacement MIC-KEY* SF Feeding Tube or conventional gastrostomy tube at home.
The life span of the balloon varies according to several factors: medication, volume of water used to
fill the balloon, gastric pH, and tube care.

FEEDING TUBE BLOCKAGE

1. Attach the MIC-KEY* SF Extension Set to the Feed Port of the MIC-KEY* SF Feeding Tube. Place
an ENFit
Syringe filled with 10–20 ml of warm water into the Extension Set Feeding/Medication
®
Port.
2. Gently pull back on, then depress the plunger to dislodge the clog.
3. If the clog remains, repeat the previous step. Gentle suction alternating with syringe pressure will
clear most obstructions. If this fails, consult your Medical Professional.
If your Medical Professional has given you different instructions for this procedure, follow them
instead of what appears here.
To prevent blockages, flush your tube:
1. Before and after each feeding.
2. Before and after giving medications.
3. Every 3 to 4 hours if the patient is receiving continuous feedings.
4. After checking for stomach content residuals.
Important: Always flush through the Extension Set. Using a syringe in the Feed Port can damage
the anti-reflux valve and cause leaking. Avoid using acidic irrigants such as cranberry juice and
carbonated soft drinks to flush, as the acidity, when combined with formula proteins, may actually
contribute to clogging.