Varian Turbo-V 150HT Mode D'emploi page 76

Masquer les pouces Voir aussi pour Turbo-V 150HT:
Table des Matières

Publicité

Press the PUMP CURRENT push-button to enter the data.
The run up time is factory-set to:
00h 08m 00s.
When the last digit is entered, the display shows:
D E L A Y
E V E N
T H R E S H O L D :
where: XXX = YES or NO.
Press the CYCLE NUMBER push-button and
YES if relay R2 must operate only after the run-up time
or select NO when the R2 operation is needed right
from start of the turbopump and after the rotational
speed of the turbopump exceeds for the first time the
speed threshold value (see the preceding cycle
diagram).
This function is factory-set to YES.
D E L A Y
E V E N
T H R E S H O L D :
A F T E R
X X X
select
A F T E R
Y E S
70
TECHNICAL INFORMATION
Press PUMP CURRENT to confirm. and the display
shows:
P U M P
L
I
R E S E T ?
where:
PUMP LIFE = is the elapsed operating time range 000
to 99999 hours.
RESET XXX = YES or NO.
The controller is factory-set to NO.
If YES is selected, the pump life shall be reset to 000.
After selecting YES, press the PUMP CURRENT push-
button to enter the command and the display shows:
R E A D Y
F O R
S O F T
S T A R T
or
P U M P
R E A D Y
S T A R T
and the controller is ready to restart (see paragraph
"Startup").
When PUMP LIFE is reset to 000, the CYCLE number
is also reset to 000.
F E
X X X X X h
X X X
L O C A L
:
P U S H
B U T T O N
NOTE
87-900-893-01 (C)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9435969-9535

Table des Matières