NVM175 Handbook
Special Instructions for medical applications
IEC/EN 60601-1 2nd Edition
UL 60601-1 1st Edition
CSA-C22.2 No. 601.1-M90
IEC/EN 60601-1 3rd Edition
ANSI/AAMI ES 60601-1
CSA 22.2 No 60601-1
These products are designed for continuous operation within an overall enclosure, and must be
mounted such that access to the mains terminals is restricted. See the appropriate standard listed
above.
For Class II installation, these products need to be fixed such that they are isolated from unearthed
accessible conductive parts by at least 2 MOPP's.
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air
or with oxygen, or with nitrous oxide.
For IEC/EN 60601-1 2nd Edition, UL 60601-1 1st Edition, CSA-C22.2 No. 601.1-M90, these products
have a reinforced insulation barrier between input and output.
For IEC/EN 60601-1 3rd Edition, ANSI/AAMI ES 60601-1, CSA 22.2 No 60601-1, these products
provide reinforced insulation between input and outputs of 2 MOPPs. 1 MOPP from input to earth and
1 MOPP from output to earth.
These products are suitable for B and BF type medical equipment.
These products are NOT protected against the ingress of water.
All outputs have basic spacings to earth rated for mains at 250Vac, and due consideration must be
given to this in the end product design.
These products have SELV outputs.
Reference should be made to local regulations concerning the disposal of these products at the end
of their useful life.
Where any part of this product is made accessible to the operator in the end use equipment, the
operator must not touch this part and the patient at the same time.
These products have not been assessed to IEC/EN60601-1-2 (EMC) but EMC test data is available
from TDK-Lambda UK Ltd.
WARNING: No modification of this product is allowed.
Environmental Specifications:
Description
Use
Temperature
Humidity
Altitude
Pressure
Orientation
Material Group
Pollution Degree
Overvoltage Category
Class
Weight
IP Rating
* 3000m for 60601-1
* 4000m for 60601-1 for model X50018A
Level of Insulation:
Dielectric Strength type testing is carried out as follows:
Primary mains circuit to earth: 2.515 to 2.525kVDC
Operation
Indoor
0°C - +70°C (See O/P tables for deratings)
5 - 95% RH, non-condensing
-200m - 5000m*
54kPa - 106kPa
The unit may be mounted on either side, vertical
with input lowest and horizontal. (Customer Air
versions can be mounted in any orientation).
Page 13 of 23
IIIb
2
II
I or II (depending on model)
1 Kg max
IPX0
Storage & Transportation
-
-40°C - +85°C
5 - 95% RH, non-condensing
-200m - 5000m
54kPa - 106kPa
All
17814 issue 17, September 2019