Recommended cleaning, packaging for sterilization, and sterilization procedures are provided below.

Cleaning

Instruments, Handpiece Connector, Combitorque
• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Remove the instruments from handpieces.
• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Treat the parts with ultrasound for at least 3 minutes.
• Submerge all components completely in deionized water.
• Rinse through all product lumens with deionized water.
• Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible on the product.
• Dry all components at room temperature.
Handpieces
• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Remove the instruments from the handpieces. Remove the handpieces from the connectors. Remove the gaskets, caps and other removable parts.
• Brush all lumens and surfaces with a suitable lumen / surface brush until all visible soiling is removed.
back rinsing), using a disposable syringe (min. volume 50 ml) applied to the nozzles of the product.
• Submerge all components completely in deionized water.
disposable syringe (min. volume 50 ml) applied to the nozzles of the product.
• Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible on the product.
• Dry all components at room temperature.

Packaging for Sterilization

The clean, dry instruments should be placed in the Steribox sterilization container. The container should be bagged in a sterilization pouch that is validated for
use in a 132ºC pre-vacuum steam sterilization cycle. The sterilization pouch should be sealed according to the manufacturer's instructions.

Sterilization

patient use. Hu-Friedy & EMS recommends sterilization in a fractional prevacuum cycle at 132°C for 4 min with a drying time of 20 minutes. Hu-Friedy & EMS
recommends use of biological indicators and chemical indicators that have been validated for use in a 132ºC pre-vacuum steam sterilization cycle.