 Reading 30 - readings in succession: Press and
release the START button 3 consecutively to recall the
last 30 readings in succession.
Pressing and releasing the START button 3 after the last
30 readings have been recalled will resume the above
sequence from reading 1.

9. Error Messages

 Measured temperature too high AR: Displays «H»
when measured temperature is higher than 42.2 °C /
108 °F in body mode or 100 °C / 212 °F in object mode.
 Measured temperature too low AS: Displays «L»
when measured temperature is lower than 34.0 °C / 93.2
°F in body mode or 0 °C / 32 °F in object mode.
 Ambient temperature too high BT: Displays «H»
and
when ambient temperature is higher than
40.0 °C / 104.0 °F.
 Ambient temperature too low BK: Displays «L» and
when ambient temperature is lower than 16.0 °C /
60.8 °F in body mode or lower than 5.0 °C / 41.0 °F in
object mode.
 Error function display BL: The system has a malfunction.
 Blank display BM: Check if the batteries have been
inserted correctly. Also check polarity (<+> and <->) of
the batteries.
6
 Flat battery indicator BN: If only «» icon is shown on
the display, the batteries should be replaced immediately.

10.Cleaning and Disinfecting

Use an alcohol swab or cotton tissue moistened with
alcohol (70% Isopropyl) to clean the thermometer casing
and the measuring sensor. Ensure that no liquid enters the
interior of the device. Never use abrasive cleaning agents,
thinners or benzene for cleaning and never immerse the
device in water or other cleaning liquids. Take care not to
scratch the surface of the sensor lens and the display.

11.Battery Replacement

This device is supplied with 2 new, long-life 1.5V, size AAA
batteries. Batteries need replacing when this icon «» BN
is the only symbol shown on the display.
Remove the battery cover BO by sliding it in the direction
shown.
Replace the batteries – ensure correct polarity as shown by
the symbols in the compartment.
Batteries and electronic devices must be dis-
posed of in accordance with the locally applicable
regulations, not with domestic waste.

12.Guarantee

This device is covered by a 2 year guarantee from the date
of purchase. During this guarantee period, at our discre-
tion, Microlife will repair or replace the defective product
free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
 Transport costs and risks of transport.
 Damage caused by incorrect application or non-compli-
ance with the instructions for use.
 Damage caused by leaking batteries.
 Damage caused by accident or misuse.
 Packaging/storage material and instructions for use.
 Regular checks and maintenance (calibration).
 Accessories and wearing parts: Battery.
Should guarantee service be required, please contact the
dealer from where the product was purchased, or your local
Microlife service. You may contact your local Microlife
service through our website:
www.microlife.com/support
Compensation is limited to the value of the product. The
guarantee will be granted if the complete product is
returned with the original invoice. Repair or replacement
within guarantee does not prolong or renew the guarantee