WARNING
To avoid possible electric shock or personal injury, follow these
guidelines:
Do not use the meter if it is damaged. Before you use the
meter, inspect the case. Pay particular attention to the
insulation surrounding the connectors
Inspect the test leads for damaged insulation or exposed
metal. Check the test leads for continuity. Replace damaged
test leads before you use the meter.
Do not use the meter if it operates abnormally. Protection
may be impaired. When in doubt, have the meter serviced.
Do not operate the meter around explosive gas, vapor, or
dust.
Do not apply more than the rated voltage, as marked on the
meter, between terminals or between any terminal and earth
ground.
Before use, verify the meter's operation by measuring a
known voltage.
When servicing the meter, use only specified replacement
parts.
Use with caution when working above 30V AC RMS, 42V
peak, or 60V dc. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger
guards on the probes.
Connect the common test lead before you connect the live
test lead. When you disconnect test leads, disconnect the live
test lead first.
Remove the test leads from the meter before you open the
battery door.
Do not operate the meter with the battery door or portions of
the cover removed or loosened.
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the batteries as soon as the
low battery indicator ("
") appears.