DC CURRENT MEASUREMENT
1)
[600.219]
Connect the black test lead to the "COM" jack and the red
test lead to the "mA" jack (While the current to be measured
is between 200MA and 20A, remove the red test lead to the
"20A" jack.)
[600.524]
Connect the black test lead to the "COM" jack and the red
test lead to the "2A" jack (While the current to be measured
is between 2A and 20A, remove the red test lead to the
"20A" jack.)
2)
Set the Function / Range switch to the desired "
the current to be measured is not known beforehand, set the
switch to the highest range and then turn down range by
range until satisfactory resolution is obtained.
3)
Open the circuit in which the current is to be measured, and
connect the test leads in series with the circuit.
4)
Read the current value displayed on the LCD.
RESISTANCE MEASUREMENT
1)
Connect the red test lead to the "V/Ω" jack and the black
test lead to the "COM" jack.
2)
Set the Function / Range switch to the desired Ω range.
3)
Connect the test leads to the resistor to be measured and
read the value displayed on the LCD.
Note:
•
For resistance about 1MΩ and above, the meter may twake
a few seconds to stabilize. This is normal for high resistance
readings.
CAPACITANCE MEASUREMENT
1)
Set the Function / Range switch to the desired Cx range.
2)
Before inserting the capacitor to be measured into the
capacitance measuring socket, be sure that the capacitor
has been fully discharged.
3)
Insert the capacitor to be measured into the capacitance
measuring socket.
4)
Read the capacitance value displayed on the LCD.
"
range. If