Span Calibration (Cont.)
Y
1

Linearity Calibration

Linearity calibration uses three calibration points; zero, mid-scale and full scale.
Lin Cal must be selected and set to YES in the .S.E.T.U.P. Menu. Before beginning
calibration, make sure the menu Lock Switch is off. Clear the platform.
Start with the balance OFF.
5
• • •
N
1
N
1
(Place mass value on platform)
(Place mass value on platform)
(Remove mass)
(Remove mass)
Y
1
Y
1
Y
1
Y
1