Maintenance
Test passed, with reservations:
This state means that partial tests of the favourite could not be
conducted due to special settings.
Confirmation is shown on the display.
The status LED flashes yellow.
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
WARNING
!
Prior to each use of the gas detector, check whether it
is switched on and in measuring mode after removing
it from the X-dock. Otherwise the user may carry a
device that is switched off!
Test not passed:
An error message is shown on the display.
The Status LED flashes red.
If required, select the desired instrument field for additional
information.
Identify and rectify the error.
Repeat the test if necessary.
24
0 0
01133286.eps
0 0
01133286.eps
Overview of status LEDs
Colour
Status
blue
flashing
green
flashing
yellow
flashing
red
flashing
5.5
After use
1. If required, remove instruments from modules.
2. Close the test gas cylinders.
NOTICE
i
i
To keep energy consumption low, Dräger recommends
switching off the station after use according to the
Instructions for Use.
If the gas monitors are stored in the station, this will
lead to increased power consumption by the gas
monitors. If the gas monitors are stored in the station,
Dräger recommends using the X-am 125+ modules.
6
Maintenance
6.1
Maintenance intervals
NOTICE
i
i
The maintenance intervals must be established in
each individual case and shortened if necessary,
depending on safety considerations, process
conditions, and the technical requirements of the
equipment. Dräger recommend a Dräger service
contract for all maintenance activities and that all
repairs are carried out by Dräger.
6.1.1
Before every start-up
The following work must be carried out before every start-up of
the equipment:
Check the hoses for dirt, brittleness and damage and
replace if necessary.
Check the hoses are secure, to prevent escapes of gas.
Check that all cable connections are secure.
Visual inspection of the modules and sensor seals. If very
dirty or if there are visible defects, the sensor seal must be
replaced.
6.1.2
Annually
Inspection of the entire X-dock station by competent
personnel.
Meaning
Process in progress
Test passed
Test passed,
with reservations:
Test failed/cancelled
Dräger X-dock 5300/6300/6600