Télécharger Imprimer la page

Dräger X-dock 5300 Notice D'utilisation page 20

Masquer les pouces Voir aussi pour X-dock 5300:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
Basics
4.5
Logging users in or out
NOTICE
i
i
A User-ID is required to log in. This must be created
beforehand by the administrator.
A user with administrator rights is created by default:
User name: admin
Password: 123456
NOTICE
i
i
Dräger recommends changing the admin password
after initial start-up.
To log in a user:
1. Select
.
a. Select
.
b. Select the desired user name from the list.
or
a. Select Select user.
b. Enter desired user name.
2. Enter the password and confirm with
NOTICE
i
i
When entering the user name, 3 already saved user
names will automatically be suggested and displayed.
For quick selection, please select the desired user
name.
To log out the current user:
1. Select
.
Information about the current user will be displayed.
2. Select
.
The current user will be logged out.
4.6
Configuring the test gas inlet
WARNING
!
The gas concentrations entered must be identical to
the specifications on the gas cylinder used. Incorrect
details will result in faulty measurement results.
NOTICE
i
i
When the test gas concentration is changed, the
corresponding test gas inlet must be reconfigured.
To configure a gas inlet:
1. Select
> Station gas configuration.
An overview of the test gas connections is displayed.
20
Select the test gas inlet
2. Select desired gas inlet.
The configuration menu will appear.
Select the test gas
inlet
Enter part no.
Enter use-by date
.
Select test gas
Enter
concentration
Further options
When using a Dräger gas cylinder:
i
i
1. Enter the part no. of the Dräger gas cylinder.
All the necessary details for the configuration will be filled
in automatically. The batch number and the expiry date can
be entered manually in addition.
i
i
2. If necessary select Further options and
gas cylinder level indicator.
3. If required, configure other gas inlets in the same way.
When using a gas cylinder from another manufacturer:
NOTICE
Upon entering the part no. of a Dräger gas cylinder, a
gas cylinder level indicator is displayed automatically,
unless this function was deactivated before (see
chapter 4.6.1 on page 21).
NOTICE
The values entered automatically must be matched
with those indicated on the test gas cylinder. If these
values are not identical, the value indicated on the gas
cylinder is to be considered and a manual correction of
the values must be made in the station.
Dräger X-dock 5300/6300/6600
◄ Enter batch no.
Create Test gas
component
◄ Delete
configuration
◄ Select
unit
, to reset the

Publicité

loading

Ce manuel est également adapté pour:

X-dock 6300X-dock 6600