ACE INSTRUMENTS ADVISES
EVEN IF YOU GET A RESULT MEANING „SAFE RANGE", YOU SHOULD NEVER DRIVE AFTER CONSUMING
ALCOHOL OR DO ANY OTHER ACTIVITIES THAT ARE SEVERELY IMPAIRED BY ALCOHOL CONSUMPTION AND
MIGHT ENDANGER YOUR HEALTH AND SAFETY OR THE HEALTH AND SAFETY OF OTHERS.
Display Indications
Display
Meaning
BAT
The testing cannot be performed because
the battery is low.
CAL
The device should be calibrated (to
maintain measurement accuracy and
sensor warranty).
Flo
The sampling volume or blow power is not
enough to analyze the breath sample.
Out
If no breath sample was given within 3
minutes, the device turns off automatically.
This also happens after 1 minute without
using the device.
SEN
The sensor is defective.
13
Action
Replace with new batteries. Be aware of the
polarity when inserting the batteries (+/- pole).
Please send the device in for calibration or
purchase a new sensor. You can send the
device for calibration to the service address
given at www.alkomat.net or order a new sensor
in the shop.
Switch the device on again and blow
steadily, constantly and long enough into
the mouthpiece or mouthpiece opening
(depending on the selected test mode). It is
best to blow as if you were whistling without
a sound.
Press power button to restart the countdown
cycle and retest.
Purchase a new sensor or send the instrument
to ACE Instruments (www.alkomat.net) for
service.
Change of the Sensor
1. Loosen the screw on
2. Carefully remove the cover and then the sensor
the top of the device but
module like it is shown on the two pictures on the
don`t pull it out.
right side. Do NOT detach the cable from the body!
4. After preparing a new
5. Insert the sensor into
calibrated or new sensor
the frame.
module insert the cable
into the new sensor. Clip
on the new sensor on
the metal blank.
Please note: Put the
blue stripe on the blue
point!
7. Re-attach the cover
8. The front of the cover has to fit exactly into the
on the top of the body.
body. If the pins in front of the sensor don`t fit,
It should snap in fairly
please carefully press them back into the body.
easily.
Calibration
To maintain the accuracy of the breathalyzer, ACE Instruments recommends a calibration after 500 tests or
every 6 to 12 months (depending which incident happens first). For a calibration service you can send the
device to ACE Instruments (see www.alkomat.net). The calibration is not mandatory, but the accuracy of the
breathalyzer will decrease steadily and additionally the warranty for the sensor expires.
We recommend sending it as a parcel, so that in case of loss an inquiry or proof of delivery is possible.
3. Remove the cable
from the slot underneath
the sensor module and
pull the old sensor off
from the metal blank.
6. The two pictures on the right show the correct
position of the sensor.
9. To finish the process,
size the screw again.