Technical Specifications
Pressure-flow
curve
Dead space
information
Therapy
pressure
Resistance
Environmental
conditions
Sound
Gross
dimensions
Mask setting
options
The mask contains passive venting to protect against rebreathing.
As a result of manufacturing variations, the vent flow rate may
vary.
Physical dead space is the empty volume of the mask to the end of
the swivel. Using the wide cushions it is 104 mL.
4 to 30 cm H
O
2
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
at 100 L/min: 1.2 cm H
Operating temperature: +5°C to +40°C
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance
with ISO 4871. The A-weighted sound power level of the mask is
31 dBA, with uncertainty of 3 dBA. The A-weighted sound pressure
level of the mask at a distance of 1 m is 23 dBA, with uncertainty of
3 dBA.
483 mm (H) x 155 mm (W) x 110 mm (D)
Mask fully assembled with short tube assembly (no headgear)
For S9: Select 'Pillows'
For other devices: Select 'SWIFT' (if available), otherwise select
'MIRAGE' as the mask option.
O
2
O
2
Pressure
Flow
(cm H
O)
(L/min)
2
4
20
8
29
11
35
12
37
16
43
17
44
20
49
22.5
53
24
54
30
61
English