Instructions for use
6567460 D3643.201.01.03.02
1.
Check whether the instrument is clean after reprocessing
under good lighting (min. 500 lux) and color rendering index
(min. 80 Ra).
2.
If they are still dirty, repeat the process.
Ä
For further reprocessing, the instrument should be dry and
free of residue.
3.
Blow the instrument out with max. 3 bar.
4.
Lubricate mechanical parts manually [→ 40].
5.
Maintain the push button chuck manually.
6.
Pack the instrument in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite
packaging according to ISO 11607.
7.
Perform sterilization [→ 44].
7 Reprocessing
37