These assembly instructions describe how to fit an alternative 12 V
plug to the following charging interface/connecting cable:
WM number
WM 22436
WM 10650
Prerequisites for an alternative 12 V plug
Suitable for a 13.8 V on-board power supply
The temperature range must correspond to the internal
temperatures arising during use
Current load > 6 A
Strain relief is essential
Remove existing 12 V plug
1. Cut connecting cable approx. 1 cm behind the 12 V plug.
2. Dispose of the plug you have cut off as electronic waste.
Attach the alternative 12 V plug
1. Remove the insulation from the cable ends.
2. Twist cable shielding and black cable together, then crimp with
a cable end-sleeve.
3. Twist red and brown cable together, then crimp with a cable
end-sleeve.