Patient Setup
1. Make sure the Pre-Use Checkout Procedure has been performed.
9
Figure 9
2. Clamp the connective tubing and turn the mode selector switch to I
(On).
Clamp
3. Set the prescribed suction level.
WARNING
⚠ The regulator must be occluded when setting the
4. Turn the mode selector switch to O (Off).
5. Attach the connective tubing to the vacuum port of the collection container.
6. Turn the mode selector switch to | (On).
Important: The Low models do not deliver suction in excess of 135 mmHg
± 5 mmHg (18.0. kPa ± 0.7 kPa).
400
500
60
70
High Flow
High Vacuum
600
80
90
100
700
prescribed suction level so that the patient does nor
receive higher than required suction.
6700-0355-000 (Rev. 9) 08/2020