Operation

An Ohio Medical overflow safety trap and high flow suction filter should be used
between the collection container and regulator to prevent contamination of the
regulator. Hospital-supplied suction tubing must be used between the catheter
and the collection bottle. Recommended minimum inside diameter is 0.25 in
(6 mm).
ISO 10079-3 (section 5.1.2) states that "the usable volume of the collection
container shall not be less than 500 ml."
High Flow Suction Filters
Hydrophilic
Pkg of 20
Pkg of 200
Hydrophobic Filter
Pkg of 3
Pkg of 10
Pkg of 100
Attaching the Overflow Safety Trap (OST)
CAUTION
⚠ To help prevent aspirate from entering the regulator, wall outlet
Trap fitting
2
Figure 2
1. Raise the sleeve and insert the trap into the regulator fitting.
2. Turn the trap clockwise about one and a half
turns to engage the threads. The trap does
not need to be screwed tight; an O-ring in
the regulator fitting provides a vacuum seal.
The trap should rotate freely to allow the
desired tubing positioning.
3. Lower sleeve to lock trap in position.
Regulator
Sleeve
6730-0350-800
6730-0351-800
Tubing x tubing nipple
6730-0570-800
6730-0571-800
6730-0572-800
and pipeline equipment, an overflow safety trap should be
attached prior to its use. Aspirate in the regulator, wall outlet
and pipeline equipment may impair their operation. The use
of the overflow safety trap and suction filter will help prevent
this and extend the life of suction equipment.
6700-0355-000 (Rev. 9) 08/2020
1/8" NPT x tubing nipple
6730-0580-800
6730-0581-800
6730-0582-800
5