Cleaning
Routine cleaning of the regulator is recommended as a standard procedure after
each use.
Wipe all exterior surfaces with a solution of water and mild detergent.
CAUTION
⚠ Do not steam autoclave or liquid sterilize the regulator. Severe
WARNINGS ⚠ Clean and disinfect all suction equipment before shipment
⚠ After patient use, regulators may be contaminated. Handle
Troubleshooting
If the regulator does not operate and you have performed the Pre-Use Checkout
Procedure, the following procedures may be used to attempt to correct the problem.
Problem
Possible Cause
No suction
Mode selector is in the O (Off)
position or between positions
Leak in system
Suction control knob at full
decrease
Important: If the above actions do not correct the problems or other problems
exist, refer to CVR service manual or refer servicing to qualified service personnel.
Warranty
This Product is sold by Ohio Medical, LLC (the "Company") under the express terms of
the warranty set forth below.
For a period of THIRTY SIX (36) MONTHS from the date the Company ships this Product to
the customer, but in no event for a period of more than three years from the date of original
delivery by the Company to an authorized dealer, this Product, other than its expendable
parts, is warranted to be free from functional defects in materials and workmanship and
to conform in all material respects to the description for the Product contained in this
10
impairment to the operation of the regulator will result.
to ensure transportation personnel and service personnel
are not exposed to any hazardous contamination.
in accordance with your hospital's infection control policy.
6700-0355-000 (Rev. 9) 08/2020
Remedy
Switch to I (On)
Check that lid is secure on
the collection container
Check tubing connections
Rotate the suction control
knob in the increase
direction (clockwise)