The Speedy is intended for use in the electromagnetic environment specified below.
The customer or the user of the Speedy should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2kV for power supply
transient/burst
lines
IEC 61000-4-4
± 1kV for input/output
lines
Surge IEC
± 1kV differential mode
61000-4-5
± 2kV common mode
Voltage Dips,
<5% UT(>95% dip in
short interrup-
UT) for 0,5 cycle
tions and voltage
40% UT(60% dip in
variations on
UT) for 5 cycles
power supply
70% UT(30% dip in
input lines IEC
UT) for 25 cycles
61000-4-11
<5% UT(>95% dip in
UT) for 5 s
Power frequency
3 A/m
(50/60 Hz) mag-
netic field IEC
61000-4-8
Recommended separation distance between portable and mobile RF communications equipment and the Speedy
The Speedy is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the Speedy can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Speedy as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of trans-
mitter / W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and manufacturer's declaration-electromagnetic immunity
Compliance
level
± 6 kV contact
± 8 kV air
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
3 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz / d=1,2√P
0,12
0,38
1,2
3,8
12
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Speedy requires continued operation
during power mains interruptions, it is recommended that the Speedy
be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
80 MHz to 800 MHz / d=1,2√P
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz / d=2,3√P
0,23
0,73
2,3
7,3
23
41