Troubleshooting
Error
Problem
Message
The system is not function-
ing properly.
Exceeding operating tem-
perature range
(1) In Temple mode: Tem-
perature taken is higher
than +42.2°C (108°F)
(2) In Object mode: Tem-
perature taken is higher
than 99°C (210.2°F)
(1) In Temple mode: Tem-
perature taken is lower
than +34°C (93.2°F)
(2) In Object mode: Tem-
perature taken is lower
than 0°C (32°F)
Device cannot be powered
on to the ready stage.
Solution
Unload the battery, wait for 1 minute and
re-power it. If the message reappears, con-
tact the retailer for service.
Allow the thermometer to rest in a room for
at least 30 minutes at room temperature:
15~40°C (59~104°F).
Please select the target within specifications.
If a malfunction still exists, please contact
the nearest retailer.
Please select the target within specifications.
If a malfunction still exists, please contact
the nearest retailer.
Change with a new battery.
23