Product specifications
This product may only be operated if its safe use is guaranteed. Observe the notes on safety and usage!
VANGUARD AG assumes no liability for damages that arise from improper operation of the Vanguard Fixed
Loop Diagnostic Catheter NAV eco BWN and the Vanguard Variable Loop Diagnostic Catheter NAV eco BWO
and/or disregard of these instructions for use.
VANGUARD AG and the manufacturer of the original product, Biosense Webster, Inc., (33 Technology Drive,
Irvine, California 92618, USA) are not affiliated companies; the device is fully refurbished solely by Vanguard
AG, who markets the device without collaboration by Biosense Webster, Inc.
"Biosense Webster" or "LASSO NAV ECO" and "LASSO 2515 NAV ECO" are protected brands of the
manufacturer of the original product Biosense Webster Inc., USA or one of its affiliated companies.
PRODUCT DESCRIPTION
The Vanguard Fixed Loop Diagnostic Catheter NAV eco BWN and the Vanguard Variable Loop Diagnostic
Catheter NAV eco BWO consists of a torsion-resistant shaft and proximally a steerable area (curvature) with a
width-adjustable loop (only the Vanguard Variable Loop Diagnostic Catheter NAV eco BWO) with several
attached electrodes. This catheter is intended for temporary intracardiac recording of potentials, stimulation of
the heart during electrophysiological examinations and for mapping of the heart's atrial structures.
The curvature of the proximal shaft area is controlled by a movable plunger with a thumb knob in the hand
piece of the catheter. To adjust the curvature, the thumb knob is pushed forward. When the thumb knob is in
the rear position, the curvature is straight. The loop width can be adjusted (only the Vanguard Variable Loop
Diagnostic Catheter NAV eco BWO ) within a certain range using a rotary knob. To constrict the loop, turn the
knob in clockwise direction, and turn it in counter-clockwise direction to expand the loop.
An extension cable with an appropriate plug connector should be used to connect it to a suitable standard
recording device.
Intended use
The Vanguard Fixed Loop Diagnostic Catheter NAV eco BWN and the Vanguard Variable Loop Diagnostic
Catheter NAV eco BWO are designed for temporary intracardial ECG recording and stimulation (mapping) of
the cardiac structures within the scope of electrophysiological examinations.
The catheter is not suitable for ablation.
The Vanguard Fixed Loop Diagnostic Catheter NAV eco BWN and the Vanguard Variable Loop Diagnostic
Catheter NAV eco BWO provides location information when used with Carto 3 EP Navigation System. It is not
compatible with Carto 3 EP Navigation System prior to Version 2.3.
It is the responsibility of the physician to use a suitable surgical procedure and technology. The procedure
described in the instructions for use is for information purposes only. Each physician must apply, supplement
or adapt the information in the instructions for use according to his/her medical training and clinical
experience. The medical device may only be used by a trained and experienced specialist physician.
Technical data
©Vanguard AG 1000001802_Rev_B_22.07.2020
12