Modulation b)
Pulse modulation b) 18 Hz
FM c) ± 5kHz deviation
1kHz sine
Pulse modulation b)
217 Hz
Pulse modulation b) 18 Hz
Pulse modulation b)
217 Hz
Pulse modulation b)
217 Hz
Pulse modulation b)
217 Hz
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK
MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced
minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS
shall be calculated using the following equation:
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the
IMMUNITY TEST LEVEL in V/m.
Modulation b) (W)
1.8
2
0.2
2
2
2
0.2
E =
Distance (m)
0.3
0.3
0.3
0.3
0.3
0.3
0.3
EMC GUIDANCE
Immunity Test Level
(V/m)
27
28
9
28
28
28
9