Soehnle Systo Monitor Connect 300 Mode D'emploi page 170

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 40
EMC GUIDANCE
Table 5
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device, should assure that it is used in such an environment.
Radiated RF IEC61000 4-3
(Test specifications for EN-
CLOSURE PORT IMMUNITY
to RF wireless communica-
tions equipment)
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test
distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while
it does not represent actual modulation, it would be worst case.
168
Test Frequency
Band a) MHz)
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5240
5785
Service a)
380-390
TETRA 400
GMRS 460,
380-390
FRS 460
704-787
LTE Band 13, 17
GSM 800/900,
TETRA 800, iDEN 820,
800-960
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
1700-1990
GSM 1900; DECT;
LTE Band 1, 3, 4, 25;
UMTS
Bluetooth, WLAN,
2400-2570
802.11 b/g/n, RFID
2450, LTE Band 7
5100-5800
WLAN 802.11 a/n

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Systo monitor 300

Table des Matières