7. Setup mode
Use the Setup Mode to change different parameters.
The thermometer stores the settings in its memory.
7.1 Setup options
Option
Type
Offset
Sleep Mode
7.2 Entering or Exiting Setup
When the thermometer is in Setup mode, the display shows SETUP.
Press
Set
button for 3 seconds to enter or exit Setup mode.
7.3 Changing a Setup Option
1. Press
HOLD
2. Press
Enter
3. Press
HOLD
display.
4. Press
Enter
Note:
Setup is disabled in MAX/MIN mode.
7.4 Offset correction
Use the offset option in Setup to adjust the thermometer's readings
to compensate for the errors of a speciic thermocouple.
The allowable adjustment range is ±5,0 °C or ±9.0 °F.
1. Plug the thermocouple into the input connector TI or T2.
2. Place the thermocouple in a known, stable temperature environment
3. Allow the readings to stabilize.
4. In Setup Mode change the offset until the top reading matches the
calibration temperature.
7.5 Auto Power Off
The thermometer will automatically switch off after 20 minutes if no
button is pressed during this time.
In Setup mode:
Press
or
HOLD
"On" or "OFF". Press
appears on the display.
Press
to store the new setting in memory.
Enter
On (sleep mode on) or 0FF (sleep mode off).
8
www.skf.indpart.ru
Menu item
TYPE
OFFSET
SLP
or
to scroll to the setup option you want to change.
MAX
MIN
to indicate that you want to change this setting.
or
until the setting you want to use appears on the
MAX
MIN
to store the new setting in memory.
to scroll to the "SLP" page. Press
MAX
MIN
or
HOLD