Reset Command; Starting The Pump - Varian Turbo-V 301 Notice De Mode D'emploi

Table des Matières

Publicité

Reset Command

The display shows:
P U M P
L
I
R E S E T ?
where:
PUMP LIFE = is the elapsed operating time range 000
to 99999 hours.
RESET XXX = YES or NO.
If YES is selected, the pump life shall be reset to 000.
After selecting YES, press the PUMP CURRENT push-
button to enter the command and the display shows as
follows.
The controller is factory-set to NO.
When PUMP LIFE is reset to 000, the CYCLE number
is also reset to 000.

Starting the Pump

If the forepump and vent device are not operated by the
controller, close the vent valve and switch on the
forepump.
With the FRONT panel operation selected, the
REMOTE and RS 232 operations are inoperative;
conversely,
the
CYCLE
CURRENT push-buttons are always active, even when
the operating mode selected is REMOTE or RS 232.
F E
X X X X X h
X X X
NOTE
NOTE
NUMBER
and
PUMP
Press the START push-button (or use the remote or
RS 232 start signal), and the display shows:
S O F T
S T A R T
[
.
.
.
Where the sign minus (-) become a square (
when
the
pump
The active step is indicated
square (
).
As the ten steps are fully covered, the pump will
reach the Normal operation. If during the Soft Start
mode the current drawn by the pump exceed 1.6 A
the speed of the pump is decreased to maintain the
maximum power allowable (1.6 A).
If within each step of the Soft Start mode the
rotational speed of the pump do not rech the
planned value within 15 minutes the display shows:
F A U L T
:
O V E R T
I M E
where X = the step number from 0 to 9 indicating
the step number not passed.
When this message is displayed the pump is
stopped.
Reset the controller by pressing twice the STOP
push-button to start the pump.
82
TECHNICAL INFORMATION
O N
.
.
.
.
.
.
.
]
finish
the
ramp-up
by a flashing
:
S X
87-900-940-01 (F)
)
step.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9437969-9537

Table des Matières