Press the CURRENT push-button and the display
shows:
I
= 0
.
0 0 A
X
.
.
K R P M
where:
I = is the DC current drawn by the pump range (0.00
to 9.99 Ampere)
P = is the DC power drawn by the pump (range 0 to
999 Watt)
KRPM = is the theoretical rotational speed of the
pump as a function of the controller output
frequency (range 9 to 75 KRPM)
°C = is the temperature of the outer ring of the
upper bearing (range 00 to 99 °C)
X = during operation a selected set point condition
(1 or 2 contrast inverted) appears when the
programmed threshold speed value is not reached.
Press the CYCLE NUMBER once and the display
shows:
R E A D Y
F O R
S O F T
S T A R T
P U M P
R E A D Y :
S T A R T
P =
0 W
X X ° C
L O C A L
P U S H
B U T T O N
Press the the CYCLE NUMBER twice and the display
shows:
X X X X C Y C L E
P U M P
L I
where:
CYCLE = are the cycles performed (range 0 to 9999)
m = is the elapsed time related to the cycle number
displayed (range 0 to 99999 minutes)
PUMP LIFE = is the total operation time of the pump
(range 0 to 99999 hours).

FRONT / REMOTE Selection

By pressing CYCLE NUMBER and PUMP CURRENT
pushbuttons together for at least 2 seconds, the
processor enters in a routine where it is possible to
program the controller. In this routine, the CYCLE
pushbutton is used for choosing/changing the value or
condition; the PUMP CURRENT pushbutton is used to
enter and confirm the value.
The display shows:
S O F T
S T A R T
S E L E C T I
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump
to ramp-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is
deselected and the ramp-up of the pump will be done
within 75 seconds.
The controller is factory set to YES.
56
TECHNICAL INFORMATION
X X X X X m
F E
X X X X X h
M O D E
O N :
X X X
87-900-871-01 (B)