GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC
IMMUNITY – FOR DEVICE THAT IS NOT LIFE SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an
environment.
IEC 60601
Immunity test
test level
3 Vrms
Conducted RF
IEC 61000-4-6
150 kHz to
80 MHz
3 V/m
Radiated RF
IEC 61000-4-3
80 MHz to
2,5 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
136
Compli-
Electromagnetic environment – guidance
ance level
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than
the recommended separation distance
3 V/m
calculated from the equation applicable to
the frequency of the transmitter
Recommended separation distance
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol: