The power indi-
If the low battery indicator lights, it
cator is on, and
means that the batteries are low or
the nebulizer
the power is insufficient to operate
does not work
the device
The membrane of the medication
chamber is damaged
The contacts electrodes on the
medication chamber are contami-
nated with medicine or water
The electrodes on the nebulizer or medi-
cation chamber are contaminated
The membrane of the medication
chamber is strongly clogged
The nebulizer
The medication chamber is not
turns off during
installed up to the stop
the use
The micro - USB cable is not prop-
erly connected to the nebulizer
Out of drugs
The drugs do not enter the
membrane for more than 15 seconds
The nebulizer was shaken when
The nebuliz-
using
ing process is
unstable
The nebulizer
The medication chamber is broken Replace the medication cham-
does not turn off
automatically
when out of the
Some solutions for inhalation can lead
drugs
to the formation of a large amount of
foam in the medication chamber
The contacts electrodes on the
medication chamber are contami-
nated with medicine or water
The electrodes on the nebulizer or
medication chamber are contami-
nated
Leakage of
Damage to the medication chamber
drugs from
or run out of the silicone gasket
the medication
chamber
If your nebulizer still does not work properly (go to agu-baby.com).
EN
Replace the batteries and reinstall
the device.
(See page 110)
Replace the medication cham-
ber with a new one and fill in the
drugs. (See page 113)
Clean the contacts electrodes from
the remnants of drugs or water and
restart the device. (See page 118)
Clean the electrodes and restart
the device. (See page 118)
If it still not work normally after clean-
ing, replace the medication chamber
with a new one.
Install the medication chamber
correctly and restart the device.
(See page 113)
Connect it correctly and restart
the device. (See page 109)
Add the drugs to the medication
chamber. (See page 113)
Place the nebulizer at right angle that
the drugs flow to the mem-brane
Hold the nebulizer securely in
your hand. (See page 115)
ber with a new one and fill in the
drugs. (See page 113)
Clean the foam and restart the
device. (See page 113)
Clean the contacts electrodes and
restart the device. (See page 118)
Clean the electrodes and restart
the device. (See page 118)
Replace the medication cham-
ber with a new one and fill in the
drugs. (See page 113)