IEC 60601 Test
Immunity test
level
6 kV contact
Electrostatic
discharge
8 kV air
IEC 61000-4-2
Fast electri-
2 kV for electri-
cal transients/
cal power lines
flashes
1 kV for input/
output lines
IEC 61000-4-4
Flashes
1 kV between
IEC 61000-4-5
the lines
2 kV between
the line (s) and
the ground
<5% UT (UT
Interruptions
and voltage
drop >95%)
variations on
during the half
the input power
cycle
lines
40% UT (UT
IEC 61000-4-11
drop 60%)
during 5 cycles
70% UT (UT
drop 30%)
during 25 cycles
<5% UT (UT
drop >95%)
during 5
seconds
Magnet fields
3 A/m
of the power
frequency
(50/60 Hz)
IEC 61000-4-8
NOTE: UT is the voltage in the supply network before the test begins
EN
Compliance
Electromagnetic environ-
level
ment-guidance
6 kV contact
The floor must be wood-
8 kV air
en, concrete or covered with
ceramic tiles. If the floor is
covered with synthetic materi-
als, the relative humidity should
be at least 30%
2 kV for elec-
The quality of electricity in the
trical power
power network must meet the
lines
standards for power supply
systems for industrial or medi-
1 kV for input/
cal environments
output lines
1 kV between
The quality of electricity in the
the lines
power network must meet the
2 kV between
standards for power supply
the line (s) and
systems for industrial or medi-
the ground
cal environments
<5% UT (UT
The quality of electricity in the
drop >95%)
power network must meet the
during the half
standards for power supply
cycle
systems for industrial or medi-
cal environments. If the user
40% UT (UT
needs the continuous operation
drop 60%)
of the device in the presence
during 5 cycles
of voltage interruptions, it is
recommended to use batteries
70% UT (UT
or an uninterruptible power
drop 30%)
supply to power the device
during 25
cycles
<5% UT (UT
drop >95%)
during 5
seconds
3 A/m
The quality of electricity in the
power network must meet the
standards for power supply
systems for industrial or medi-
cal environments
131