Lithium battery replacement by inadequately trained personnel could result in a hazard such as a fire or
explosion.
Do not plug or unplug the power cord into the electrical outlet with wet hands. If the AC adapter is abnormal,
please change the adapter.
Do not pull out the adapter when you are using the monitor.
Do not use any other type of AC adapter as it may harm the monitor.
The monitor, cable, battery and cuff must be disposed of according to local regulations at the end of their usage.
Note: Battery life and charge cycles vary by use and settings.

TROUBLESHOOTING

PROBLEM
Low Battery
LED display reads"Er0"
LED display reads"Er1"
LED display reads"Er2"
LED display reads"Er3"
LED display reads"Er4"
POSSIBLE CAUSE
Battery do not have enough power
Pressure system is unstable before measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff is too
tight during inflation
Pneumatic system leakage or cuff is too
loose during inflation
SOLUTION
Charge the battery
Retest, make sure not to move
your arm or the monitor
Apply the cuff correctly and try again
10