Handling during measurement
Press the operating lever (C) of the caliper arm completely downwards and bring the
gage in measuring position.
Please be careful with the measuring contacts (A1+B1) and the movable caliper
arm (A) and try to manage the gage in measuring position without touching the
workpiece.
Hold the gage loose during measurement and release the operating lever (C).
To determine the correct measuring value, please pivot or displace the gage and read
from the display the minimum or maximum deflection of the pointer.
The gage software is available with different features and options which allows an easy
measurement. For more information see Chapter "Measuring programs".
Please check the instrument via comparison measurement with a gage block
each time before starting a new measurement series ( zero point
adjustment).
The instrument starts the first time with "Absolute – Mode" and "standard
program".
When starting the instrument the last mode will be shown.
The gage switches back into the last used measuring mode, if no key has been
pushed for more than 30s in the setup menu. Changes in the setup menu will
be adopted.
Display
Follow up constantly the measured value shown in the display.
When a reverse point is found within a range of ±20 digits the pointer and the suitable
analog bar will be shown.
If a new reversal point of more than ±20 digits is found the analog bar will be adjusted
automatically.
(This program will be started during the initial use or after factory setting.)
7