3 MEASURING PRESSURE
The measuring pressure is obtained by means
of the compression spring 7 (Fig. 5). In standard
version, the measuring pressure is 2 N at electro-
nic zero. Other measuring pressures can be ob-
tained by changing the "Bearer/spring" assembly 7
(Fig. 5), the colour of which indicates the nominal
pressure.
Springs are available for the following pressures:
4 MOUNTING POSITION
The probe may be mounted in any position. If it
is mounted horizontally (Fig. 2 A) and B)), mea-
suring pressure "F" is the nominal force of the
spring which is fitted. If it is mounted vertically
0,4 N
Red
0,63 N
Yellow
1,0 N
Green
1,6 N
Blue
2,5 N
Brown
4,0 N
Black
2,0 N
Nickeled
(Fig. 2 C) and D)), the measuring pressure is set
when the device is mounted by changing and
ajusting the spring beacause the weight of the
mobile elements may vary.
FMS
03260448
03260449
03260450
03260451
03260452
03260453
Standard
spring
7