To turn Process Enforced Usage ON, OFF or to activate DOCUMENTATION mode, select SETTINGS then USER and
follow these steps:
Scroll to PROCESS and select.
1.
2.
Select ENFORCED.
3.
Use the arrows to select one of the following:
ON, OFF or DOCUMENTATION.
Press the BACK icon to save your change and return to
4.
the main menu.
Any user can stop a cycle and remove the load even with the Process Enforced feature ON. However, the cycle
data will record that an unregistered user has stopped the cycle and/or opened the door.
Using Process Enforced Documentation mode:
This mode activates the process enforced usage function along with the additional documentation function that generates a
report with information about the cycle and the load type.
When starting a cycle with Process Enforced Documentation
mode turned ON, you will be prompted to identify the general
contents of the load to be processed by selecting from a list as
well as whether a biological indicator and chemical indicator
are included.
41