Hach LZY904.98.00002 Instructions D'utilisation page 19

Masquer les pouces Voir aussi pour LZY904.98.00002:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 71
9. Install the calibration
lid. Make sure that the
calibration lid is in the
closed position. Refer
to
Installation
on page 3.
13. Remove the vial.
17. Install the process
head (or the automatic
cleaning module).
Section 4 Verification
Do a calibration verification immediately after each calibration to measure the verification standard
and record the measured value to the instrument.
10. If the standard
value that shows on the
display is not correct,
enter the accurate
turbidity value of the
standard from the
certificate of analysis.
If the standard value
that shows on the
display is correct, push
enter.
14. Do steps 4 to 12
again until all of the
standard vials are
measured.
18. Push ENTER to
save the calibration
value. The status
indicator light stays
green.
11. Complete the steps
that show on the
controller display.
15. If the value of the
verification standard
shows on the display,
do steps 6 to 12 again
to measure the
verification standard.
12. When the status
indicator light changes
to green, remove the
calibration lid.
16. Make sure that
there is no water on the
process head (or the
automatic cleaning
module). Dry all
possible spills to
prevent water ingress
on the vial
compartment.
English 19