Hach LZY904.98.00002 Instructions D'utilisation page 10

Masquer les pouces Voir aussi pour LZY904.98.00002:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 71
9. When the status
indicator light stays
blue, remove the
calibration lid.
13. Install the
calibration lid. Make
sure that the calibration
lid is in the closed
position. Refer to
Installation
on page 3.
17. Remove the vial.
10 English
10. Remove the vial.
14. Push the button on
the front of the
instrument.
18. Make sure that
there is no water on the
process head (or the
automatic cleaning
module). Dry all
possible spills to
prevent water ingress
on the vial
compartment.
11. Put the verification
standard vial in front of
the RFID module. A
beep sound is heard
and the status indicator
light flashes blue. If the
status indicator light
does not flashes blue
refer to
Troubleshooting
on page 22.
The instrument records
the value, the lot
number, the expiration
date and the Certificate
of Analysis information
from the RFID vial to
the data log.
15. Wait 15 to
20 seconds for the
measurement to
complete. The status
indicator light slowly
flashes blue during the
measurement.
19. Hold the process
head (or automatic
cleaning module)
vertically when it is
installed on the
instrument or the vial
can break.
12. Put the verification
standard vial in the vial
compartment.
16. When the status
indicator light flashes
green, remove the
calibration lid.
20. Push the button on
the front of the
instrument to save the
calibration value. The
status indicator light
stays green.

Publicité

Table des Matières
loading

Table des Matières