LEEP PRECISION
System Description
The CooperSurgical LEEP PRECISION Smoke Evacuation System
LP-30-220 three-stage air filtration system is used to remove airborne
particulate plumes produced during office and surgical procedures and
has the following features:
• Low noise level.
• Triple filtration of air provides efficiency level for 0.014 microns rated
at 99.999 percent. This includes a pre-filter, a charcoal filter for odor
removal and a final safety filter placed after the charcoal filter.
• High air flow for effective plume collection.
• Conveniently attaches to the CooperSurgical LEEP PRECISION
Integrated System—wall-mounted or free-standing.
Cautions
• U.S. Federal law restricts this device to sale by or on the order
of a physician.
• This device produces a strong vacuum force and care should be
exercised to ensure that the suction and the position of the inlet end
of the suction tubing is adjusted properly to prevent injury to the
patient or inadvertent damage to surgical materials.
• The materials in the plume removed by this device are potentially
hazardous. Handle according to 29 CFR 1910.1030 and OSHA
3127.1992 (Occupational Exposure to Bloodborne Pathogens)
guidelines.
• To prevent a fire or explosion hazard, do not use the system in the
presence of flammable or potentially flammable materials.
• Do not allow fluid to be pulled into the system.
• To prevent premature failure of the ULPA filter cylinder, do not
operate this device without a disposable Pre-Filter in place.
• No modification of this equipment is allowed.
• No Customer serviceable parts.
• Only replacement of power cords or fuses can be done by the user.
Only replace with these items with the exact replacement part(s)
available from CooperSurgical.
Warning
To avoid risk of Electrical Shock, this equipment must only be connected
to a supply mains with protected earth.
TM