Treatment is
Power wire is
uncomfort-
worn or malfunc-
able
tioning
Electrodes are
worn or malfunc-
tioning
Intensity level set
too high
Electrodes are not
wet enough
Irregular
Power wire mal-
stimulation
functioning
Stimulation
Electrodes pos-
is ineffect-
itioned incorrectly
ive
Replace the power
wire
Replace the cuff
Lower intensity
level
Shut off the de-
vice and remoisten
electrodes
Lower intensity
level and rotate the
power wire plug
90°. If problem
persists, replace
the power wire
Remove elec-
trodes and repos-
ition, ensuring cuff
has sufficient skin
contact
11
The skin
The universal cuff
turns red
is not positioned
and/or a
correctly on the
stabbing
skin
pain occurs
The universal cuff
is dirty
One of the water
contact elec-
trodes has a
scratch
Water contact
electrodes are not
sufficiently moist
Ensure that the
universal cuff sits
comfortably on the
skin and cannot
move
Clean the universal
cuff as described
in these ins- truc-
tions for use
Replace the uni-
versal cuff
Switch off the de-
vice. Moisten elec-
trodes