Table 3. For EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
Infrared thermometer AGU IHE3 is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of Infrared thermometer AGU IHE3
should assure that it is used in such environment
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: at 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
EN
IEC 60601
Compliance
test level
level
3 Vrms
3 Vrms
150 kHz
to 80 MHz
3 V/m
3 V/m
80 MHz
to 2.5 GHz
Electromagnetic environment
guidance
Portable and mobile
RF communications equipment
should be used no closer to
any part of AGU IHE3 infrared
thermometer, including cables,
than the recommended sepa-
ration distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d=1.2√P,
d=1.2√P 80 MHz to 800 MHz,
d=2.3√P 800 MHz to 2.5 GHz,
where P is the maximum
output power rating of the
transmitter in watts (W) accord-
ing to the transmitter manu-
facturer and d is the recom-
mended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
should be less than the
compliance level in each
frequency range
. Interference
b
may occur in the vicinity of
equipment marked with the
following symbol:
a