11
TROUBLESHOOTING
If a problem occurs during operation, first of all, refer to the list of possible fail-
ures given below.
Failure
The temperature measured is
The temperature measured is
The operating temperature
STORAGE, TRANSPORTATION
12
AND OPERATION REQUIREMENTS
• The device shall be transported within the temperature range from
–10 °С to +40 °С with relative humidity up to 75 %.
• The device can be operated within the temperature range from +15 °С to
+35 °С with relative humidity up to 15–95 %.
• The device shall be stored within the temperature range from +10 °С to
+30 °С with relative humidity up to 65 %.
• Do not expose to thermal shock.
ATTENTION
After transporting or storing at low temperatures, it is necessary to keep
the device at room temperature for at least 2 hours before switching on.
EN
Failure description
higher than 43 °C (109.4 °F)
lower than 34 °C (93.2 °F)
is not in the range 15~35 °C
(59~95 °F)
Remedial action
Operate the thermometer only
between the specified temperature
ranges. If necessary, clean the
sensor tip. In the event of a repeated
error message, contact your retailer
or Customer Service
Operate the device only
at an ambient temperature within
the specified values
63