6. Care and Maintenance
The unit will pump all the waste water from the sump. If a small amount remains, soak it
up with absorbent material. Wear disposable rubber gloves. Dispose of absorbent material
according to biological waste disposal regulations.
If the unit is without power, use the drain tube located under the front of the unit in the
middle to drain the water from the chamber and system. Then unlock the door using the
door lever (see section 4.9) and remove the coarse filter and fine filter (see section 6.4) and
use absorbent material to soak up any water remaining in the sump.

6.7 Annual maintenance

If the message "Preventative Maintenance Required. Call for Service" appears, the annual
maintenance is required. To continue using the HYDRIM, press OK. Call for service as soon
as possible. The annual maintenance should be done after 365 days have elapsed since the
initial installation or since the last service call was performed.

6.8 Unit testing and validation

The following outlines the recommended regular maintenance to be performed by the user
and technicians for the HYDRIM C61w G4 washer.
To ensure the unit's ongoing cleaning efficacy, SciCan recommends observing the following
periodic testing protocols.
The recommended periodic testing protocols are listed in the table below:
Upon installation
Daily tests and checks
Annually
Bi-Annually
When
SciCan-approved
SciCan-approved
SciCan-approved
Who
Commissioning/Installation testing as
By a
outlined in the installation document included
with the unit.
technician
First validation if required by local regulations.
door lock check
wash arm rotation check
door seal check
By the user
load carrier check
check and clean chamber filters
visual examination (inspection under
magnification) of each load for residual soil.
By a
Annual maintenance schedule
as outlined in the service manual
technician
By a
Re-validation if required by local regulations.
technician