9. Turn hex wrench clockwise to engage locking mechanism. Once the wrench clicks, the anchor is
locked to lead.
10. If repositioning of the lead is required after the anchor has been secured to the lead, insert hex wrench
into set screw opening and turn counterclockwise to loosen set screw.
CAUTION: Do not excessively loosen the set screw. This may cause the set screw to dislodge.
OPTIONAL: Reinsert stylet into lead
CAUTION: If it is difficult to reinsert the stylet, move the anchor to ensure that the set screw has been
loosened sufficiently. Do not use excessive force when reinserting the stylet into the leads.
Once set screw is loosened, the anchor may be moved along the lead. Ensure that sutures have also
been removed to allow for anchor to be repositioned. Once anchor is repositioned, proceed to Step 7.