Calculations for nominal volume V
N
x
= results of weighings
i
n = number of weighings
Z = correction factor
(e. g., 1.0029 µl/mg at 20 °C, 1013 hPa)
Mean value
Accuracy
Σ x
V – V
A% =
· 100
x =
i
N
n
V
N
Mean volume
Coefficient of variation
100 s
V = x · Z
CV% =
V
Standard deviation
Σ (x
– x )
2
i
s = Z ·
n – 1
33