7.3. Adjusting the volume

(only piccolo 2)
Warning!
Reagent will dispense while volume is being adjusted. Hold a suitable
vessel below the discharge tube outlet.
To set the volume, push down the
operating knob and twist handle 90°. The
nose of the housing indicates the volume
after releasing operating knob.

7.4. Dispensing

1. Place a suitable vessel below the discharge tube.
2. Slowly steadily depress piston, release gently. No air bubbles should
be visible during dispensing.
Warning!
If the piston moves stiffy or is ifficult to move, stop dispensing
immediately and clean the instrument (see chap. 8).
Attention!
Reagent may escape from the discharge tube.
1.
2.
3.
27