DC AND AC CURRENT MEASUREMENT (See Fig. 3)
❶ Connect red test lead to the A input (ACx for HD115B) for current measurements up to 2A or to the 10A
input for current measurements to 10A. Connect black test lead to COM input connector. ❷ Set the Function
Switch to the desired current range. ❸ Press the AC/DC menu button to display either AC or DC indicator. ❹
Open circuit in which current is to be measured (voltage between this point and ground must not exceed
1000V). Securely connect test leads in series with the load. š Turn on power to circuit being tested. › Read
current value on Digital Display.
Note: If, in the 10A range, the meter is exposed to current greater than 10A, it should be turned off and
allowed to cool for at least 10 minutes
Incorrect Input Warning: A tone will sound when a test lead is connected to either amperage jack, but the
selector switch is not set to the correct current range. Ranges 200mA to 2A require the test lead in the A jack
(ACx for HD115B). The 10A range is only for use with the test lead in the 10A jack.
4a
Discharge capacitors
4b
Kondensatoren entladen
Descargue los condensadores
Décharger les condensateurs
4c
5
4d
Fig. 3
A
or
6
AC
HD110B
480V
3
240V
AC
120V
50V
24V
OFF
1500
1000
SAFETY
TM
TESTER
500V AC
MAX
2
200u
2m
20m
200m
A
2A
10A
10A
A
2A MAX
10A MAX
FUSED
FUSED
1
- 14 -
A
mA
DC
DATA
HOLD
200
20
V
2
200m
20M
2M
200k
20k
2k
200
V
COM
CAT III
MAX
1500V
1000V
red/rot/
roja/rouge

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Hd115b

Table des Matières