Troubleshooting
CAUTION
Any oil leaking from the X-ray tube in the event of a fault is harmful.
Wipe up any oil immediately�
i
Do not swallow the oil�
i
Stop using the unit and inform a service technician�
i
11 Tips for operators and service technicians
Any repairs above and beyond routine maintenance may only be done by suitably qualified
personnel or by one of our service technicians�
Fault
Unit does not switch on
Unit not responding
2210200496L41 1904V004
OBJ_DOKU-0000003575-mehrere-005.indd 43
Probable cause
EMERGENCY STOP SWITCH ac-
cidentally activated
No mains voltage
On/off switch is defective
The unit has not yet completed
the startup procedure
Cables not correctly connected
Plug-in contacts of the fiber optic
cable are contaminated
Driver for PCI Express frame
grabber card is not installed or in-
correctly installed
COM port incorrectly configured
Solution
Releasing the EMERGENCY
i
STOP SWITCH�
Check the mains cable and
i
electrical connection; replace if
necessary�
Contact technician�
i
Check the mains fuse in the
i
building�
Contact technician�
i
After switching on, wait until the
i
booting process has finished�
Check the cable connections�
i
Clean plug-in contacts and
i
sockets�
Install driver or complete
i
VistaVox Plugin once again�
Check the COM port in the
i
service tool�